The Slimline system on average over-recorded esophageal acid exposure compared to the Bravo system largely because of a flawed software scheme for electrode thermal calibration. Accuracy of pH data sets from both systems can be improved by scrutiny for artifacts and use of an in vivo pH reference.
No takes yet. Share an insight, caveat, or question.
Pandolfino et al. (2005) studied this question.