Key result
Custom analog circuitry matches the MAX30102 sensor for basic HRV metrics with improved signal-to-noise ratio.
Why the study?
PPG is commonly used for wearable HRV monitoring, motivating a comparison between a standard sensor and custom analog circuitry.
Does a custom analog circuitry with an ADS1115 ADC provide comparable PPG-derived HRV metrics to a MAX30102 sensor?
Does a custom analog circuitry with an ADS1115 ADC provide comparable PPG-derived HRV metrics to a MAX30102 sensor?
A custom analog acquisition system with an ADS1115 ADC can reliably reproduce basic HRV parameters from PPG signals with improved SNR compared to the MAX30102 sensor, offering a low-cost biosensing alternative.
Supports broader clinical use of affordable PPG for routine HRV tracking; extends validation to low-cost sensors versus custom high-resolution ADC.
Heart rate variability (HRV) is a key physiological marker for autonomic nervous system function and cardiovascular health. Photoplethysmography (PPG) is commonly used to derive HRV metrics in wearable and low-cost monitoring systems. This study presents a comparative assessment of basic HRV metrics obtained from a MAX30102 optical sensor and a custom analog circuitry with an ADS1115 analog-to-digital converter (ADC). Both measurement pathways were carefully aligned using analog high-pass and low-pass filters and a consistent digital filtering pipeline, ensuring that the frequency bands relevant to HRV were preserved. PPG signals were recorded simultaneously, and inter-beat intervals were extracted to calculate the Standard Deviation of NN intervals (SDNN), Root Mean Square of Successive Differences (RMSSD), and Percentage of successive NN intervals >50 ms (pNN50) across multiple 30-s windows. Bland–Altman analysis was employed to evaluate agreement between the two methods. Results indicate that the analog circuit with an ADS1115 achieves comparable HRV basic metrics to the MAX30102 sensor, with improved Signal-to-Noise Ratio (SNR) due to high-resolution ADC and low-noise analog amplification. These findings demonstrate that a carefully designed analog acquisition system can reliably reproduce HRV basic parameters from PPG signals, providing an alternative approach for low-cost, flexible biosensing platforms.
No takes yet. Share an insight, caveat, or question.
Miranda-Vega et al. (2026) studied this question. Custom analog circuitry with ADS1115 ADC vs. MAX30102 optical sensor was evaluated on Agreement in basic HRV metrics (SDNN, RMSSD, pNN50) and Signal-to-Noise Ratio. A custom analog circuitry with an ADS1115 ADC achieved comparable basic HRV metrics to the MAX30102 sensor with improved signal-to-noise ratio.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: