Key result
A mean absolute deviation (MAD) score ≤12% between RVOT VT and a pace-map was 93% sensitive and 75% specific for identifying a successful ablation site (9.5% vs 13.3%; p=0.01).
Why the study?
Comparison of 12-lead ECG morphology between ventricular tachycardia and a pace-map is frequently performed, but objective criteria for quantifying waveform similarity are lacking.
Does quantitative comparison of 12-lead ECG waveforms using mean absolute deviation (MAD) identify successful ablation sites in patients with right ventricular outflow tract ventricular tachycardia?
Observational (n=15)
Does quantitative comparison of 12-lead ECG waveforms using mean absolute deviation (MAD) identify successful ablation sites in patients with right ventricular outflow tract ventricular tachycardia?
Absolute Event Rate: 9.5% vs 13.3%
p-value: p=0.01
A mean absolute deviation (MAD) score <=12% between RVOT VT and a pace-map provides an objective, highly sensitive metric for identifying successful ablation sites.
May aid objective pace-mapping in RVOT VT ablation; leaves open prospective validation before clinical adoption.
OBJECTIVES: The purpose of this study was to objectively quantify the similarity of 12-lead electrocardiogram (ECG) waveforms using two quantitative metrics, the correlation coefficient (CORR) and the mean absolute deviation (MAD). BACKGROUND: Comparison of the 12-lead ECG morphology between ventricular tachycardia (VT) and a pace-map is frequently performed; however, there are no objective criteria for quantifying the similarity between two waveform morphologies. METHODS: During ablation of right ventricular outflow tract (RVOT) VT, 12-lead ECG pace-maps were acquired from three superior septal sites, three superior free wall sites, and before each ablation attempt in 15 patients. The 12-lead ECG waveforms of the clinical tachycardia and pace-maps were compared using both MAD and CORR at each site. RESULTS: The MAD scores were lower (i.e., more closely matched) for septal compared with free wall sites (15.9 +/- 5.3% vs. 25.3 +/- 10.2%; p < 0.001). Successful ablation sites had a significantly lower MAD score compared with unsuccessful sites (9.5 +/- 2.8% vs. 13.3 +/- 5.6%; p = 0.01), whereas there was only a trend toward a higher CORR for successful ablation sites (98.2 +/- 1.2% vs. 96 +/- 4.7%; p = 0.07). A MAD score < or =12% was 93% sensitive and 75% specific for identifying a successful ablation site. There was an inverse correlation between MAD score and distance from the site of VT origin (r = 0.63, p < 0.001). CONCLUSIONS: A MAD score >12% between RVOT VT and a pace-map at any site suggests sufficient dissimilarity to dissuade ablation at that site. The MAD score can be used to standardize 12-lead ECG waveform morphology comparisons among different laboratories, and may be useful for guiding ablation of VT.
No takes yet. Share an insight, caveat, or question.
Gerstenfeld et al. (2003) conducted an observational in Right ventricular outflow tract ventricular tachycardia (n=15). Mean absolute deviation (MAD) and correlation coefficient (CORR) of 12-lead ECG waveforms vs. Successful vs unsuccessful ablation sites was evaluated on MAD score for successful vs unsuccessful ablation sites (p=0.01). A mean absolute deviation (MAD) score ≤12% between RVOT VT and a pace-map was 93% sensitive and 75% specific for identifying a successful ablation site (9.5% vs 13.3%; p=0.01).