Case report highlights ventilation failure risk in infants using fenestrated tracheostomy tubes, suggesting caution.
Pediatric tracheostomy is a common procedure, with a substantial proportion of cases performed on infants, often due to chronic respiratory failure or complex airway anomalies. This case report details a critical intraoperative event involving a three-month-old infant undergoing tracheostomy for a laryngeal mass. The patient experienced immediate and profound ventilation failure following the insertion of a fenestrated tracheostomy tube. Mechanical obstruction was ruled out by passing a suction catheter through the tube. Furthermore, switching ventilatory modes to pressure-controlled volume-guaranteed (PC-VG) or volume-controlled ventilation (VCV) was also ineffective in restoring adequate ventilation. This was managed using a completely unfenestrated tube. The tube was placed and the breathing circuit connected, a large portion of the air leak was successfully eliminated, and appropriate ventilation parameters were restored. Fenestrated tracheostomy tubes have established benefits in adults; however, this case highlights a cautionary observation in ventilated infants due to potential air leak and ventilatory failure, underscoring careful tube selection.
No takes yet. Share an insight, caveat, or question.
Hazaymeh et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: