Key result
Improved spatial vectorcardiography matches 3D torso-model accuracy while correcting for torso shape and avoiding left-arm leads.
Why the study?
Clinical use of spatial vectorcardiography required an improved system offering an optimum compromise among theoretic soundness, accuracy, reproducibility, signal-to-noise ratio, and speed of application.
Introduces a new, clinically practical system for spatial vectorcardiography with improved accuracy and physical basis.
Answers are grounded in this paper and cite their sources.
May enhance vectorcardiography accuracy; leaves open effects on clinical outcomes.
This paper describes a news improved system of spatial vectorcardiography that is practical for clinical use. It represents an optimum compromise among such factors as soundness of theoretic basis, accuracy, reproducibility, signal-to-noise ratio, and speed of application. Some of its advantages over currently employed systems include a rational physical basis, corrections for torso shape, avoidance of left arm, insensitivity to individual variability of ventricle location, and accuracy comparable to applicability of 3-dimensional torso-model data to the human subject. Detailed description of electrode placement, practical procedures, and useful technics is included.
No takes yet. Share an insight, caveat, or question.
Ernest Frank (1956) studied this question. Spatial vectorcardiography system vs. Currently employed vectorcardiography systems was evaluated. An improved spatial vectorcardiography system offers torso shape corrections, left arm avoidance, and accuracy comparable to 3-dimensional torso-model data.
Synapse has enriched 10 related papers. Consider them for comparative context: