Small (mm size) distal electrodes of active medical implants can cause localized radio frequency (RF)‐induced power deposition with extreme spatial gradient (5–6 dB/mm). We propose a method that aims to improve the accuracy of the traditional experimental evaluation, which relies solely on hardware instrumentation. A numerically derived local deposition distribution is used to assist the estimation of power deposition. The method was successfully validated against numerical simulations, considering three different generic implants with a single electrode. The power deposition in the surrounding region of the implant electrodes estimated with the proposed method was validated with less than 1 dB (30%) of uncertainty.