Demonstrates accurate quantification of microplastics in water, suggesting an efficient monitoring method.
Reliable quantification of microplastics in water remains challenging because most Raman-based methods require filtration, drying, or complex flow systems, which can lead to particle loss and signal instability. Here, we propose a simple dual-laser Raman strategy for the direct, real-time quantification of microplastics in water without pretreatment. By simultaneously integrating backscattering and transmission geometries using two identical 532 nm lasers, spatial variations in Raman scattering cross-sections, arising from particle motion and focal depth fluctuations, are effectively mitigated. The dual-laser configuration enhances Raman intensity by approximately 1.5-fold compared with backscattering and threefold compared with transmission alone (p < 0.001), enabling robust real-time detection with a temporal resolution of 0.1 s. Accurate particle counting is demonstrated using polystyrene (PS) standard beads and further validated for polyamide 6 (PA6) and polyvinyl chloride (PVC) particles with irregular morphologies and broad size distributions, with no false-positive events observed. By prioritizing simplicity and quantitative reliability over ultimate size resolution, the proposed strategy provides a practical approach for routine monitoring of microplastics in drinking water and industrial aqueous systems.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: