Key result
Fixed pulse width shocks linked to 100% first-shock success vs. fixed tilt for spontaneous VT/VF.
Why the study?
Shock programming has typically relied on fixed tilt waveforms, but patient-specific fixed pulse widths may improve ICD shock success for ventricular arrhythmias.
Does fixed pulse width programming improve first-shock success rates compared to 65% fixed tilt in patients with ICD/CRT-D devices treating spontaneous VT/VF?
Observational (n=277)
Yes
Does fixed pulse width programming improve first-shock success rates compared to 65% fixed tilt in patients with ICD/CRT-D devices treating spontaneous VT/VF?
Answers are grounded in this paper and cite their sources.
Programming ICDs to a fixed pulse width rather than a 65% fixed tilt significantly improves first-shock success rates for spontaneous VT/VF without increasing delivered energy.
BACKGROUND: Each phase of a biphasic implantable cardioverter-defibrillator (ICD) shock waveform can be characterized by either the percentage of peak voltage that decays (tilt) or the time taken for that decay (pulse width). Although shock programming has typically relied on fixing the tilt, patient-specific fixed pulse widths may improve shock success. OBJECTIVE: The purpose of this study was to compare shock energies and success rates to treat spontaneous ventricular tachycardia/ventricular fibrillation using ICD/CRT-D (cardiac resynchronization therapy - defibrillator) devices with fixed pulse width vs fixed tilt waveforms. METHODS: US patients with single-coil ICD or CRT-D devices (Ellipse, Assura, Unify, and Fortify, Abbott) were identified who had received defibrillation or cardioversion shock therapy (≥1 episode) when programmed to both 65% fixed tilt and fixed pulse width modes for distinct periods. Patient demographic characteristics, programming, and shock episodes were downloaded from the Merlin.net Patient Care Network (Abbott). Shock energies and success rates were compared across shock waveform types within each patient. RESULTS: In total, 277 patients experiencing at least 1 shock-delivered episode with each shock type were evaluated (mean age 70 ± 14 years; 157 [57%] with ICDs; 120 [43%] with CRT-Ds). The first-shock success rate was significantly greater with fixed pulse width vs fixed tilt (median [25th, 75th percentile] 100.0% [75.0%, 100.0%] of episodes vs 87.5% [42.1%, 100.0%] of episodes; P < .001), and more patients experienced 100% first-shock success with fixed pulse width vs fixed tilt (69.3% of episodes vs 46.2% of patients; P < .001). With fixed pulse width, 79.4% of patients (220 of 277) experienced greater-or-equal first-shock success, and they did so by delivering comparable energies (33.0 [29.6, 35.1] J vs 32.9 [25.7, 36.9] J; P = .53). CONCLUSION: Reprogramming shock waveforms from 65% fixed tilt to fixed pulse width significantly improved intrapatient shock success.
No takes yet. Share an insight, caveat, or question.
Katcher et al. (2026) conducted an observational in Spontaneous ventricular tachycardia/ventricular fibrillation (n=277). Fixed pulse width shock waveform vs. 65% fixed tilt shock waveform was evaluated on First-shock success rate (p=<.001). Fixed pulse width shock waveforms improved median first-shock success rates compared to 65% fixed tilt waveforms for spontaneous VT/VF.