Why the study?
Automatic detection by cardiac devices suboptimally groups AF, flutter, or AT together as high rate events, which may delay or misdirect therapy.
Does deep learning accurately classify atrial fibrillation from atrial tachycardia using intracardiac electrograms compared to traditional classifiers?
Does deep learning accurately classify atrial fibrillation from atrial tachycardia using intracardiac electrograms compared to traditional classifiers?
Deep learning applied to intracardiac electrograms can accurately differentiate atrial fibrillation from atrial tachycardia, outperforming traditional rate- or regularity-based classifiers.
Answers are grounded in this paper and cite their sources.
May improve intracardiac AF vs AT differentiation; leaves open prospective validation before clinical adoption.
BACKGROUND: Automatic detection of atrial fibrillation (AF) by cardiac devices is increasingly common yet suboptimally groups AF, flutter or tachycardia (AT) together as 'high rate events'. This may delay or misdirect therapy. OBJECTIVE: We hypothesized that deep learning (DL) can accurately classify AF from AT by revealing electrogram (EGM) signatures. METHODS: We studied 86 patients in whom the diagnosis of AF or AT was established at electrophysiological study (25 female, 65 ± 11 years). Custom DL architectures were trained to identify AF using N = 29,340 unipolar and N = 23,760 bipolar EGM segments. We compared DL to traditional classifiers based on rate or regularity. We explained DL using computer models to assess the impact of controlled variations in shape, rate and timing on AF/AT classification in 246,067 EGMs reconstructed from clinical data. RESULTS: DL identified AF with AUC of 0.97 ± 0.04 (unipolar) and 0.92 ± 0.09 (bipolar). Rule-based classifiers misclassified ∼10-12% of cases. DL classification was explained by regularity in EGM shape (13%) or timing (26%), and rate (60%; p < 0.001), and also by a set of unipolar EGM shapes that classified as AF independent of rate or regularity. Overall, the optimal AF 'fingerprint' comprised these specific EGM shapes, >15% timing variation, <0.48 correlation in beat-to-beat EGM shapes and CL < 190 ms (p < 0.001). CONCLUSIONS: Deep learning of intracardiac EGMs can identify AF or AT via signatures of rate, regularity in timing or shape, and specific EGM shapes. Future work should examine if these signatures differ between different clinical subpopulations with AF.
No takes yet. Share an insight, caveat, or question.
Rodrigo et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: