Key result
Amplatzer Cardiac Plug closure leads to large device thrombus requiring surgical removal at 90 days.
Case Report (n=1)
This case highlights the risk of significant thrombus formation on the Amplatzer Cardiac Plug after left atrial appendage closure despite dual antiplatelet therapy.
Answers are grounded in this paper and cite their sources.
Emphasizes need for follow-up imaging after LAA closure; leaves open optimal antithrombotic regimen to prevent device-related thrombus.
A 51-year-old female underwent a percutaneous closure of the left atrial appendage with a 28-mm Amplatzer Cardiac Plug (ACP) (St Jude Medical). When she was 8 years old, she underwent a surgical closure of an atrial septal defect. She developed paroxysmal atrial fibrillation at the age of 32. When she was 48 years old, she had an ischaemic cerebellar stroke. Since then, she has been treated with coumadin. At routine follow–up, cerebral MRI microbleeds were observed, constituting a neurological contraindication for coumadin treatment. Subsequently, she was scheduled for a percutaneous left atrial appendage closure. This procedure was done in November 2011 without any complication and resulting in a correct and stable position of the device (Panels 1 and 2). She was treated with aspirin 80 mg daily and clopidrogel 75 mg daily. There were no complications and she was discharged the day after the procedure. Our patient has shown good medical compliance. At the 90-day follow-up, a control transoesophageal echocardiography was done. A very large and mobile thrombus on the connector pin of the device was seen (Panel 3). Surgical removal of the left atrial appendage including the closure device was performed 2 days later. A large mobile thrombus was seen on the connector pin of the disc (Panel 4). Laboratory results did not show any thrombofilic risk factors.
No takes yet. Share an insight, caveat, or question.
Lammers et al. (2012) conducted a case report in Paroxysmal atrial fibrillation with contraindication to coumadin (n=1). Left atrial appendage closure with an Amplatzer Cardiac Plug was evaluated. Left atrial appendage closure using an Amplatzer Cardiac Plug was complicated by a large thrombus on the device at 90 days, requiring surgical removal in a 51-year-old female.
Synapse has enriched 10 related papers. Consider them for comparative context: