Why the study?
The study evaluated the reproducibility of CMR-derived biventricular global longitudinal strain assessment across multiple centers using travelling volunteers.
Does magnetic field strength (3.0T vs. 1.5T) or temporal resolution (25 vs. 30 cardiac phases) affect the inter-site reproducibility of feature tracking-derived biventricular global longitudinal strain in healthy travelling volunteers?
Does magnetic field strength (3.0T vs. 1.5T) or temporal resolution (25 vs. 30 cardiac phases) affect the inter-site reproducibility of feature tracking-derived biventricular global longitudinal strain in healthy travelling volunteers?
Multicenter cardiovascular magnetic resonance studies assessing feature tracking-derived global longitudinal strain must standardize temporal resolution (cardiac phases) for left ventricular measurements and magnetic field strength for right ventricular measurements to ensure inter-site reproducibility.
Answers are grounded in this paper and cite their sources.
Acquisition parameters require harmonization for reproducible biventricular GLS in multicenter CMR; leaves open effects on clinical outcomes.
PURPOSE: To evaluate the reproducibility of biventricular global longitudinal strain (GLS) assessment using cardiovascular magnetic resonance in a multicenter study of travelling volunteers. METHODS: Twenty travelling volunteers were prospectively scanned at four sites with same-vendor scanners at 3.0T (sites I, II, III) and 1.5T (site IV). Cine imaging in three long-axis views was performed using a segmented balanced steady-state free precession sequence with 30 cardiac phases except site II with 25 phases. RESULTS: Imaging and post-processing were carried out successfully for 18 volunteers in a core lab setting. Pairwise comparisons revealed significant differences in left ventricular (LV) GLS between sites I and II (p < 0.001) and sites II and IV (p = 0.013), as well as in right ventricular (RV) GLS between sites I and IV (p = 0.027). RV GLS values were significantly higher at 3.0T (p = 0.024), whereas field strength had no significant impact on LV GLS (p = 0.153). Conversely, the use of 25 cardiac phases at site II was associated with significantly lower LV GLS values (p < 0.001), while RV GLS remained unaffected (p = 0.825). CONCLUSION: When applying feature tracking-based strain in a multicenter study, careful consideration should be given to the temporal resolution for LV longitudinal strain and to magnetic field strength for RV longitudinal strain.
No takes yet. Share an insight, caveat, or question.
Ammann et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: