Key result
Rotational atherectomy creates a smooth lumen ~1.2-fold larger than the burr size.
Why the study?
The effects of high-speed rotational atherectomy on lumen size and plaque morphology in obstructive atherosclerotic coronary artery disease were not fully characterized by intravascular ultrasound.
What is the effect of rotational atherectomy on lumen geometry and plaque morphology as assessed by intravascular ultrasound in obstructive atherosclerotic coronary artery disease?
Observational (n=28)
What is the effect of rotational atherectomy on lumen geometry and plaque morphology as assessed by intravascular ultrasound in obstructive atherosclerotic coronary artery disease?
Intravascular ultrasound demonstrates that rotational atherectomy effectively ablates heavily calcified plaque, creating a smooth, cylindrical lumen larger than the burr size.
Answers are grounded in this paper and cite their sources.
Supports expected lumen gain with rotational atherectomy; leaves open effects on long-term patency and clinical outcomes.
BACKGROUND: High-speed rotational atherectomy uses a diamond-coated, elliptical burr to abrade occlusive atherosclerosis, especially noncompliant calcified plaque. METHODS AND RESULTS: Intravascular ultrasound (IVUS) was used to analyze 28 patients after atherectomy. Arteries treated and imaged were left main (three), left anterior descending (12), left circumflex (five), right coronary (seven), and saphenous vein graft (one). Twenty patients had adjunct balloon angioplasty. Twenty-two (79%) target lesions were calcified; the intimal arc of calcium was 160 +/- 126 degrees (range, 0-360 degrees). After atherectomy, the intima-lumen interface was unusually distinct and circular. The lumen was larger than the largest burr used for both stand-alone (1.19 +/- 0.19-fold the largest burr size) and adjunct balloon procedures (1.30 +/- 0.15-fold the largest burr). Three-dimensional reconstruction of the ultrasound images showed a smooth lumen, especially in calcified plaque. Deviations from cylindrical geometry occurred only in areas of soft plaque or superficial tissue disruption of calcified plaque. Five patients were studied before and after rotational atherectomy. IVUS showed an increase in lumen size, a decrease in plaque-plus-media area and in arc of target lesion calcification, and no change in target lesion external elastic membrane cross-sectional area. CONCLUSIONS: Rotational atherectomy causes atheroablation with only moderate evidence of barotrauma in heavily calcified arteries, even after adjunct balloon angioplasty. The lumen is cylindrical, especially in areas of calcified plaque, and somewhat larger than the largest burr tip used.
No takes yet. Share an insight, caveat, or question.
Mintz et al. (1992) conducted an observational in obstructive atherosclerotic coronary artery disease (n=28). Rotational atherectomy was evaluated. Rotational atherectomy achieved atheroablation with the resulting lumen being 1.19 +/- 0.19-fold the largest burr size for stand-alone procedures, leaving a smooth cylindrical lumen.