Three-dimensional (3D) printing is increasingly used for the fabrication of definitive crowns; however, whether specific post-curing hardware is mandatory for clinical success remains a practical concern. This study provided a practical comparison evaluating the effect of two post-curing units on the biaxial flexural strength (BFS), Weibull modulus (m), Martens hardness (HM), indentation modulus (EIT), water sorption (WSP), and water solubility (WSL) of 3D-printed resins for permanent crowns, compared with a conventional resin composite. A total of 200 specimens were fabricated from two 3D-printed resins (Permanent Crown™ and CrownTec™) and a conventional resin composite (Filtek Universal Restorative™) used as a control. The 3D-printed specimens were post-cured using either a Formcure or an Otoflash G171 unit. WSP and WSL were measured after 90 days of water ageing, while BFS, HM, and EIT were evaluated after 24 h of storage using standardised methods. All materials exhibited WSP and WSL values within ISO limits, with the control group showing significantly higher values and superior mechanical properties. Among the 3D-printed resins, post-curing significantly affected only HM and EIT for Permanent Crown™ resin, with no significant differences in BFS. Overall, the tested 3D-printed resins demonstrated high processing stability across different curing protocols, suggesting that clinical performance remains consistent regardless of the post-curing unit used.