Key result
Impella 5.5 support during CABG caused a transient accordion phenomenon of the right subclavian artery mimicking structural injury, which resolved completely after device removal.
Why the study?
Although the perioperative benefits of Impella 5.5 in high-risk patients undergoing CABG have been described, reports on vascular complications remain limited.
Case Report (n=1)
This case highlights that Impella 5.5 insertion can cause a reversible accordion phenomenon in the subclavian artery, mimicking structural vascular injury but resolving upon device removal.
Calls for vigilance on vascular complications with Impella 5.5 in high-risk CABG; leaves open incidence and mitigation in hemodialysis patients.
Background The Impella 5.5 has recently been introduced as a high-flow mechanical circulatory support device and is increasingly used to stabilize high-risk patients undergoing coronary artery bypass grafting (CABG). Although its perioperative benefits have been described, reports on vascular complications remain limited. Case Summary A 73-year-old man on maintenance hemodialysis with progressive left ventricular dysfunction was scheduled for CABG under Impella 5.5 support. Preoperative CT showed a minimum luminal diameter of 7 mm in the right axillary artery, and the device was inserted through a prosthetic graft the day before surgery. Immediately after device placement, right radial pulsatility markedly decreased. Angiography revealed faint distal flow and an accordion-like deformation of the right subclavian artery. Because significant vascular injury was considered unlikely and the hand remained well perfused, CABG was performed without vascular reconstruction. After successful CABG, the Impella was removed on postoperative day 1, resulting in immediate restoration of right radial pressure. Postoperative CT confirmed the absence of vascular injury, and the patient recovered without any neurological or ischemic sequelae. Discussion This case represents a rare instance of transient upper-limb hypoperfusion caused by an accordion phenomenon of the subclavian artery during Impella 5.5 support. The phenomenon mimicked structural injury but resolved completely after device removal. The case underscores the importance of detailed pre-procedural vascular imaging, continuous limb perfusion monitoring, and awareness of reversible pseudo-lesions induced by Impella catheter-related straightening of tortuous vessels.
No takes yet. Share an insight, caveat, or question.
Matsui et al. (2026) conducted a case report in Progressive left ventricular dysfunction requiring CABG (n=1). Impella 5.5 was evaluated on Vascular complications (accordion phenomenon of the right subclavian artery). Impella 5.5 support during CABG caused a transient accordion phenomenon of the right subclavian artery mimicking structural injury, which resolved completely after device removal.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: