Evidence regarding the safety and efficacy of TAVR with the Venus A-Valve system remains limited for patients with pure native aortic regurgitation.
Does transfemoral transcatheter aortic valve replacement with the Venus A-Valve system improve outcomes in patients with pure native aortic regurgitation?
Does transfemoral transcatheter aortic valve replacement with the Venus A-Valve system improve outcomes in patients with pure native aortic regurgitation?
Transfemoral TAVR using the Venus A-Valve system is a feasible and safe treatment option for patients with pure native aortic regurgitation, demonstrating high procedural success and favorable one-year outcomes.
Answers are grounded in this paper and cite their sources.
Supports Venus A-Valve TAVR feasibility in pure native AR; leaves open need for RCTs to confirm outcomes versus alternatives.
BACKGROUND: Evidence about safety and efficacy of transcatheter aortic valve replacement (TAVR) with the Venus A-Valve system (Venus Medtech, Hangzhou, China) remains limited for patients with pure native aortic regurgitation (PNAR). OBJECTIVES: The single-center study sought to report the one-year clinical outcomes of the Venus A-Valve in the treatment of PNAR. METHODS: This study was a retrospective analysis of prospectively collected data. Data was from all consecutive patients who had PNAR and underwent TAVR with the Venus A-Valve system at our center from July 2020 and June 2021. Procedural and clinical outcomes up to one year were analyzed using Valve Academic Research Consortium-2 criteria. RESULTS: A total of 45 consecutive patients with PNAR underwent transfemoral TAVR with the Venus A-Valve system. The Mean age was 73.5 ± 5.5 years and 26.7% were female. All the TAVR procedures were performed via transfemoral access. Implantations were successful in 44 cases (97.8%). Only one patient was converted to surgical aortic valve replacement. No patient died intraoperatively. No second valve was implanted. In-hospital mortality rate was 2.3%. The one-year all-cause mortality rate was 4.7% without cardiovascular related death. No patient had moderate or severe paravalvular leakage during follow-up. At one year, the mean pressure gradient was 8.8 ± 0.9 mmHg, and left ventricular ejection fraction increased to 61.5 ± 3.6%. CONCLUSIONS: This single-center study demonstrated the safety and efficacy of transfemoral TAVR with the Venus A-Valve in the treatment of patients with PNAR.
No takes yet. Share an insight, caveat, or question.
Zheng et al. (2023) studied this question.