Repurposing the Gore Thoracic Branch Endoprosthesis side branch component achieved technical success and symptom resolution in three patients with complex aortic pathology unsuitable for open repair.
Case Report (n=3)
No
Does repurposing the Gore Thoracic Branch Endoprosthesis side branch component provide technical success and symptom resolution in patients with complex aortic pathologies?
The Gore Thoracic Branch Endoprosthesis side branch component can be successfully repurposed for complex aortic pathologies, including thoracic aortic stenosis and anastomotic aneurysms.
We report a novel use of the Gore Thoracic Branch Endoprosthesis (TBE) side branch component for treating complex aortic pathologies.Two patients with high-grade mid thoracic aortic stenosis underwent repair using the TBE side branch as a tapered stent-graft combined with distal balloon-expandable stenting to coarctation anatomy.A third patient with a large anastomotic aneurysm of a visceral bypass underwent physician-modified endograft repair incorporating the TBE side branch as an antegrade branch into a dilated Dacron surgical graft.All repairs were technically successful, with symptom resolution and durable short-term outcomes.
Fereydooni et al. (2026) conducted a case report in Complex aortic pathology (n=3). Gore Thoracic Branch Endoprosthesis (TBE) side branch component was evaluated on Technical success and symptom resolution. Repurposing the Gore Thoracic Branch Endoprosthesis side branch component achieved technical success and symptom resolution in three patients with complex aortic pathology unsuitable for open repair.