Assesses blood flow in vivo with a portable DCS system, suggesting improved accessibility for monitoring.
Diffuse correlation spectroscopy (DCS) is a powerful optical technique for non-invasive quantification of deep-tissue microvascular blood flow quantification. Conventional DCS systems are bulky and rely on external computing devices, limiting their point-of-care applicability. In this study, we implemented the proposed DCS system on a Xilinx Zynq-7000 system-on-chip (SoC) via hardware-software co-design approach. The proposed approach employed the programmable logic (PL) for a multi-tau digital autocorrelator and the processing system (PS) for real-time blood flow index (BFI) extraction via the Nelder-Mead algorithm. Arterial occlusion and breath-holding in vivo tests demonstrated strong consistency and high correlation with a conventional PC-based DCS system. Additionally, the proposed system significantly reduces instrumentation size and cost, enabling portable assessment of tissue microcirculation.
No takes yet. Share an insight, caveat, or question.
Fu et al. (2026) studied this question.