Answers are grounded in this paper and cite their sources.
, vs BV/TV) was then determined. The results showed that using a threshold set by the collagen background had a profound effect on the histomorphometry bone parameters when assessed by μCT. However, the differences between air and water were not significant, suggesting that comparable data can be produced in a laboratory environment when scanning porous bone samples under either wet or dry conditions- counter to common belief. Determining which is more suitable, air or water, in laboratory and in clinical μCT imaging is important to improve the quality and relevance of biomechanics research. The data with collagen as the threshold were illuminating as they showed that remodelling rates and the relative organic to mineral presence varied with BV/TV, concurring with some other recent studies [2,3,4].
No takes yet. Share an insight, caveat, or question.
Adams et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: