Answers are grounded in this paper and cite their sources.
A convenient and accurate reversed-phase high-performance liquid chromatography (RP-HPLC) method for angiotensin I-converting enzyme inhibitory peptides assay was described in this paper. The mobile phase consisted of 70% A (0.05% TFA and 0.05% triethylamine in water, pH = 2.9-3.3) and 30% B (100% acetonitrile) using an Isogradient program. The flow rate was 0.5 mL/min. The absorb wavelength was 226.5 nm; the column temperature was controlled at 25°C. This method for angiotensin I-converting enzyme inhibitory peptides assay was convenient for the Iso-gradient program. The accuracy of the RT-HPLC method was verified by analyzing ACE inhibitory activity of the hydrolysate peptides of silkworm pupae protein, and the results showed that the RT-HPLC method was available for exploring new source of angiotensin I-converting enzyme inhibitory peptides rapidly and veraciously.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: