Experimental study demonstrates sensitive non-enzymatic glucose detection using a multi-component alloy electrode, highlighting potential for durable electrochemical biosensors.
Extensive research focuses on developing cost‐effective, high‐performance electrochemical non‐enzyme glucose sensors, particularly using nanocomposites of metal oxides and multi‐component alloys. In this paper, a batch of nano‐flaky Cu x O decorated CuCoNiFeSi (Cu x O@Cu/Fe) and CuCoNiSi (Cu x O@Cu) electrode was developed by melt‐spinning following by anodization. According to the results, the sensitivity of Cu x O@Cu/Fe‐25/35 for glucose detection in the lower (0–1 mM) and higher region (1–15 and 15–21 mM) is 2.544, 1.51 and 0.64 mA mM −1 cm −2 , respectively. The limit of detection (LOD) was estimated to be 0.88 μM (signal noise ratio ( S / N ) = 3). Furthermore, the Cu x O@Cu/Fe‐25/35 possessed various merits, such as excellent selectivity, good reusability, acceptable reproducibility, satisfying long‐term stability, and high tolerance to chloride ions. The enhanced performances of the Cu x O@Cu/Fe‐25/35 electrodes are attributable to the feather‐like Cu x O and synergistic effect between Cu x O and Ni(Fe, Co) 3 Si 2 . The work provides a new pathway for detecting all glucose density by electrochemical non‐enzymatic glucose sensor.
No takes yet. Share an insight, caveat, or question.
Wen et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: