Why the study?
Does the use of dual-chamber ICDs compared to single-chamber ICDs increase the risk of periprocedural complications in patients receiving primary prevention ICDs without a pacing indication?
Does the use of dual-chamber ICDs compared to single-chamber ICDs increase the risk of periprocedural complications in patients receiving primary prevention ICDs without a pacing indication?
Routine implantation of dual-chamber ICDs in patients without a pacing indication is associated with an increased risk of periprocedural complications without evidence of clinical benefit.
Answers are grounded in this paper and cite their sources.
Cautions against routine dual-chamber ICDs without pacing indication; leaves open confirmation via randomized trials.
<0.001) in the dual-chamber ICD group. Multivariable analyses confirmed atrial lead addition as an independent predictor for "any complications" (odds ratio [OR], 1.1 [95% CI, 1.0-1.2]), for pneumo/hemothorax (odds ratio, 1.1 [95% CI, 1.0-1.4]), and for lead dislodgement (odds ratio, 1.3 [95% CI, 1.1-1.6]). Conclusions Despite lack of evidence for clinical benefit, dual-chamber ICDs are implanted for primary prevention of sudden cardiac death in a majority of patients who do not have pacing indication. This practice is associated with increased risk of periprocedural complications. Avoidance of routine implantation of atrial leads will likely improve safety outcomes.
No takes yet. Share an insight, caveat, or question.
Margolis et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: