BACKGROUND: For the forensic aim, a sensitive and specific method using headspace gas chromatography coupled with mass spectrometry (GC/MS) has been developed for the quantitative determination of ethanol in blood using n-propanol as internal standard. GC was performed in isothermal mode with a GC run-time of 5.0 min. METHODS: The quantification was performed using selected ions monitoring mode adopting a quantitative ion and qualifier ion for ethanol and the internal standard. RESULTS: The method was linear (r(2) = 0.999, in the concentration range of 39.5-1,262.9 μg/ml), specific, sensitive (limit of quantification and limit of detection of 39.5 and 0.4 μg/ml, respectively), and robust. A slightly modified method was also developed for the quantification of 50 commonly abused drunken in blood. The method used an isothermal GC program with a run-time of 5.0 min. The quantification was performed using selected ions monitoring mode and integrating the area under the peak using n-propanol as an internal standard. The method was linear 40-1,263 μg/ml and sensitive. CONCLUSIONS: The method was proved superior in speed and selectivity to previously reported methods and was successfully applied to the pharmacokinetic study of ethanol.
No takes yet. Share an insight, caveat, or question.
Xiao et al. (2014) studied this question.