Prospective study demonstrates high diagnostic accuracy of ultrasonic flow ratio in intermediate left main stenosis, supporting image-based functional evaluation.
BACKGROUND: The recently introduced ultrasonic flow ratio (UFR), is a novel fast computational method to derive fractional flow reserve (FFR) from intravascular ultrasound (IVUS) images. In the present study, we evaluate the diagnostic performance of UFR in patients with intermediate left main (LM) stenosis. METHODS: This is a prospective, single center study enrolling consecutive patients with presence of intermediated LM lesions (diameter stenosis of 30%-80% by visual estimation) underwent IVUS and FFR measurement. An independent core laboratory assessed offline UFR and IVUS-derived minimal lumen area (MLA) in a blinded fashion. RESULTS: = 0.09). Patient level diagnostic accuracy, sensitivity and specificity for UFR to identify FFR ≤ 0.80 was 82.9% (95% CI: 70.2-95.7), 93.1% (95% CI: 82.2-100.0), 58.3% (95% CI: 26.3-90.4), respectively. CONCLUSION: In patients with intermediate LM diseases, UFR was proved to be associated with acceptable correlation and high accuracy with pressure wire-based FFR as standard reference. The present study supports the use of UFR for functional evaluation of intermediate LM stenosis.