This case report describes the novel application of the FARAWAVE pulsed field ablation catheter for treating premature ventricular contractions from the right ventricular outflow tract.
The first approved ablation device for pulsed field ablation (PFA, FARAWAVE) consists of a 12-Fr catheter with five splines carrying four electrodes each that is navigated over-the-wire via a steerable sheath. Its shape can be changed from a basket-like pose to a flower-like pose depending on the underlying anatomy. In this case, a 48-year-old female with symptomatic premature ventricular contractions (PVCs) was referred for ablation. PVC morphology was compatible with an origin in the right ventricular outflow tract (RVOT). Conventional mapping with a diagnostic catheter confirmed RVOT site of origin. A steerable sheath (FARADRIVE) was advanced to the right atrium. Using a 6F pigtail catheter a guidewire was deployed in the left pulmonary artery. The 31-mm FARAWAVE ablation catheter was deployed in the RVOT in a basket like pose. Pulsed field ablation was carried out using pulses of 2.5 s duration with a voltage of 1.8 kV.