Comparative reliability study reveals that uncontrolled probe compression skews muscle thickness and stiffness, suggesting external force sensors improve measurement precision.
< 0.05) different in the ultrasound PCC condition, compared to results obtained by both examiners. The abdominal oblique external and internal muscle thickness measurements were underestimated and all LAM shear moduli were overestimated during measurements without controlling the probe compression by an external sensor. The intra-class correlation coefficient was excellent in all conditions, but the smallest detectable differences were approximately 43-60% lower during the measurements collected in PCC condition. Differences in LAM measurements between PCC and 'on-hand' conditions may be clinically irrelevant when the force applied by the probe is consciously controlled by the examiner. However, during ultrasound measurements of the LAM morphology, the potential under/over estimation should always be considered when measurements are performed without controlling probe compression by an external sensor.