Does multiple electrode aggregometry accurately detect platelet aggregation abnormalities compared to light transmission aggregometry in patients with mucocutaneous bleeding?
Does multiple electrode aggregometry accurately detect platelet aggregation abnormalities compared to light transmission aggregometry in patients with mucocutaneous bleeding?
Multiple electrode aggregometry has high negative predictive value compared to light transmission aggregometry, making it a potentially useful screening tool to rule out platelet aggregation abnormalities without requiring further LTA testing.
This is the authors' abstract. We don't add key points for this paper.
INTRODUCTION: analyzer (Roche Diagnostics) requires smaller blood volumes and less sample manipulation than LTA, making it an attractive clinical testing option. Direct comparisons of MEA with LTA for diagnosis of platelet aggregation abnormalities are few. METHODS: Ninety-nine patients (66 F/33 M; median age 26 [range 2-86] years), referred for initial laboratory evaluation of mucocutaneous bleeding, had parallel MEA/LTA testing. Concentrations of ADP, arachidonic acid (AA), collagen, and thrombin receptor-activating peptide (TRAP) that produced threshold responses in normal controls were used for testing patients. RESULTS: Twenty-nine of the 99 patients (30%) had at least one abnormal agonist response by LTA; 15 of these patients had >1 abnormal agonist response. Thirty-six patients (36%) had at least one abnormal agonist response by MEA; 27 had >1 abnormal agonist response. Sensitivity/specificity of MEA relative to LTA: ADP, 0.70/0.72; AA, 0.71/0.85; collagen, 0.85/0.71; TRAP 0.25/0.84. Negative predictive values (NPVs) for MEA relative to LTA: ADP, 0.90; AA, 0.93; collagen, 0.97; TRAP, 0.96. CONCLUSIONS: Specific abnormal results of MEA testing did not adequately predict specific abnormalities in LTA testing using threshold agonist concentrations. However, favorable NPVs suggest that MEA may be useful in screening patients for platelet aggregation abnormalities; those with normal MEA results not requiring further diagnostic testing by LTA.
Answers are grounded in this paper and cite their sources.
MEA's high NPV supports screening to exclude aggregation defects in bleeding; leaves open prospective validation vs LTA.