OBJECTIVE:: To evaluate the effects of virtual monochromatic imaging (VMI) using dual-layer spectral detector CT on the image quality of coronary CT angiography (CCTA) acquired by using a low contrast material (CM) dose. METHODS:: ) to scan 30 controls without renal insufficiency. Quantitative parameters, including CT attenuation, image noise, and contrast-to-noise ratio (CNR), were measured. The visual image quality factors of contrast enhancement, image noise, beam-hardening artefact, vessel sharpness, and overall image quality were scored on a 4-point scale. RESULTS:: The mean CT attenuation of the ascending aorta was significantly higher for 50 keV VMI than for 120 kVp. Image noise was significantly lower under the 50 keV VMI. CNR and the mean visual score for contrast enhancement were significantly higher for 50 keV VMI. There were no significant differences in the other visual image quality parameters between the two protocols. CONCLUSION:: Dual-layer spectral detector CT using 50 keV VMI enabled reducing the CM dose by 50 % without CCAT image quality degradation in patients with renal insufficiency. ADVANCES IN KNOWLEDGE:: ) can improve the diagnostic image quality of CCTA.
No takes yet. Share an insight, caveat, or question.
Oda et al. (2018) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: