Are totally implantable venous power ports safe and effective for high-pressure injections during arterial phase CT scans compared to classic peripheral cannulas?
Are totally implantable venous power ports safe and effective for high-pressure injections during arterial phase CT scans compared to classic peripheral cannulas?
Totally implantable venous power ports are safe and effective for high-pressure injections during CT scans, providing comparable arterial contrast to peripheral cannulas, though forearm ports carry a higher risk of thrombosis than chest ports.
May favor chest over forearm ports to limit thrombosis; leaves open need for RCTs before routine arterial-phase CT use.
OBJECTIVES: To evaluate the technical success, clinical outcome and safety of percutaneously placed totally implantable venous power ports (TIVPPs) approved for high-pressure injections, and to analyse their value for arterial phase CT scans. METHODS: Retrospectively, we identified 204 patients who underwent TIVPP implantation in the forearm (n=152) or chest (n=52) between November 2009 and May 2011. Implantation via an upper arm (forearm port, FP) or subclavian vein (chest port, CP) was performed under sonographic and fluoroscopic guidance. Complications were evaluated following the standards of the Society of Interventional Radiology. Power injections via TIVPPs were analysed, focusing on adequate functioning and catheter's tip location after injection. Feasibility of automatic bolus triggering, peak injection pressure and arterial phase aortic enhancement were evaluated and compared with 50 patients who had had power injections via classic peripheral cannulas. RESULTS: Technical success was 100%. Procedure-related complications were not observed. Catheter-related thrombosis was diagnosed in 15 of 152 FPs (9.9%, 0.02/100 catheter days) and in 1 of 52 CPs (1.9%, 0.002/100 catheter days) (p<0.05). Infectious complications were diagnosed in 9 of 152 FPs (5.9%, 0.014/100 catheter days) and in 2 of 52 CPs (3.8%, 0.003/100 catheter days) (p>0.05). Arterial bolus triggering succeeded in all attempts; the mean injection pressure was 213.8 psi. Aortic enhancement did not significantly differ between injections via cannulas and TIVPPs (p>0.05). CONCLUSIONS: TIVPPs can be implanted with high technical success rates, and are associated with low rates of complications if implanted with sonographic and fluoroscopic guidance. Power injections via TIVPPs are safe and result in satisfying arterial contrast. Conventional ports should be replaced by TIVPPs.
No takes yet. Share an insight, caveat, or question.
Goltz et al. (2012) studied this question.
Synapse has enriched 2 closely related papers on similar clinical questions. Consider them for comparative context: