Answers are grounded in this paper and cite their sources.
A new PCR kit (AMPLICOR CT/NG; Roche Diagnostic Systems, Inc., Branchburg, N.J.) was used as a screening tool for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae in first-void urine (FVU) specimens from 3,340 asymptomatic women attending European health care units for contraceptive advice or pregnancy termination. All samples were kept frozen (-20 degrees C) prior to testing. Chlamydia-positive samples were retested once by the plasmid-based PCR kit and also by a major outer membrane protein (MOMP) primer-based PCR. Discrepancies were resolved by using the direct immunofluorescence test (DIF) with the centrifuged sediment of the FVU specimens. Samples positive for N. gonorrhoeae were retested by chromosomal primer-based PCR and verified by a 16S RNA PCR. Of the samples tested, 1.8% were considered inhibitory by using the internal amplification control. Of 81 samples positive for C. trachomatis, 74 samples were positive by both plasmid- and MOMP-based PCRs, 6 samples were positive by plasmid-based PCR and DIF, and one sample was positive by both MOMP-based PCR and DIF. Nine samples (0.3%) were positive for N. gonorrhoeae by the chromosomal primer-based PCR; however, none of the results could be confirmed. The test offers the unique ability to identify inhibition of amplification with the optional internal control.
No takes yet. Share an insight, caveat, or question.
Bassiri et al. (1997) studied this question.