We present a highly sensitive room-temperature atomic magnetometer (AM), designed for use in biomedical applications. The magnetometer sensor head is only 2 × 2 × 5 cm3 and is constructed using readily available, low-cost optical components. The magnetic field resolution of the AM is <10 fT Hz–1/2, which is comparable to cryogenically cooled superconducting quantum interference device (SQUID) magnetometers. We present side-by-side comparisons between our AM and a SQUID magnetometer, and show that equally high quality magnetoencephalography and magnetocardiography recordings can be obtained using our AM.
No takes yet. Share an insight, caveat, or question.
Shah et al. (2013) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: