Why the study?
Flow stasis in the neo-sinus is a recognized risk factor for transcatheter heart valve thrombosis, but the impact of valve expansion at the native annulus and implant depth on flow stasis required investigation.
Does transcatheter heart valve under-expansion and implant depth reduce neo-sinus flow stasis in an in vitro model?
Does transcatheter heart valve under-expansion and implant depth reduce neo-sinus flow stasis in an in vitro model?
In an in vitro model, 10% under-expansion of transcatheter heart valves improved neo-sinus washout, suggesting a potential strategy to reduce flow stasis and thrombosis risk.
Answers are grounded in this paper and cite their sources.
May reduce neo-sinus stasis risk in TAVR; hypothesis-generating and should not yet change practice.
OBJECTIVES: We investigated the impact of (transcatheter heart valve) THV expansion at the level of the native annulus and implant depth on valve performance and neo-sinus flow stasis. BACKGROUND: Flow stasis in the neo-sinus is one of the identified risk factors of THV thrombosis. METHODS: A 29 mm CoreValve and 26 mm SAPIEN 3 were deployed under different expansions (CoreValve, SAPIEN 3) and implant depths (CoreValve) within a patient-derived aortic root in a pulse duplicator. Fluorescent dye was injected during diastole into the neo-sinus and imaged over 20 cardiac cycles. Washout times were computed as a measure of flow stasis for each deployment. RESULTS: The 10% CoreValve under-expansion improved neo-sinus washout over full expansion by 8% (p < .001), and higher CoreValve implant depth improved neo-sinus washout (p < .001). The 10% SAPIEN 3 under-expansion improved neo-sinus washout by 23% (p < .001). Under-expansion of both valve types caused higher pressure gradients and smaller effective orifice areas than full expansion. CONCLUSIONS: Neo-sinus flow stasis is influenced by THV expansion and implant depth (CoreValve). The 10% valve under-deployment (oversizing) may facilitate reduced flow stasis in the neo-sinus with minimal increase in pressure gradients. This strategy may be helpful for patient anatomies, which are in-between transcatheter valve sizes.
No takes yet. Share an insight, caveat, or question.
Madukauwa‐David et al. (2019) studied this question.