BACKGROUND: The recently launched high-throughput assays for detecting antibodies against SARS-CoV-2 has contributed to the managing strategies for the COVID-19 pandemic. This study aimed to investigate the performance of three high-throughput assays and one rapid lateral flow test relative to regulatory authorities' recommended criteria. METHODS: A total of 315 samples, including 150 pre-pandemic samples, 152 samples from SARS-CoV-2 RT-PCR positive individuals and 13 potentially cross-reactive samples were analysed with SARS-CoV-2 IgG (Abbott, Abbott Park, IL), Elecsys Anti-SARS-CoV-2 (Roche, Solna, Sweden), LIAISON SARS-CoV-2 S1/S2 IgG (DiaSorin, Saluggia, Italy) and 2019-nCOV IgG/IgM Rapid Test (Dynamiker Biotechnology Co., Tianjin, China). RESULTS: All assays performed with a high level of specificity ranging from 96.7% to 99.3%. Sensitivity differed more between the assays, Roche exhibiting the highest sensitivity of 98.7%. The corresponding figures for Abbott, DiaSorin and Dynamiker Biotechnology were 80.9%, 89.0% and 72.4%, respectively. CONCLUSIONS: The results of the evaluated SARS-CoV-2 assays vary considerably, as well as their ability to fulfil the performance criteria proposed by regulatory authorities. Introduction into clinical use in low-prevalent settings, should, therefore, be made with caution.
No takes yet. Share an insight, caveat, or question.
Ekelund et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: