LiF/ZnS scintillation screen and a silicon photomultiplier array linked to a capacitive network to detect the positions of incident neutrons, is constructed and tested. To evaluate the detector performance, a series of neutron beam experiments with the detector prototype were performed in the BL20 at the China Spallation Neutron Source. The spatial resolution was measured, and the energy-selective neutron imaging and Bragg edge measurements of a 316L stainless steel sample were performed. A sub-millimeter spatial resolution could be obtained for the detector prototype under study. The detector with such a high spatial resolution is promising for applications in neutron scattering experimental installations, especially for neutron single-crystal diffractometers.