Demonstrates increased field of view with lower element count in ultrasound imaging, suggesting improved efficiency.
Ultrafast three-dimensional (3-D) ultrasound imaging is an emerging topic in the field of medical ultrasound because it provides a full volume view of a region for applications such as obstetrics, abdominal, or cardiovascular imaging. However, ultrasound matrix arrays have an extremely high element count, e.g., 1024 elements for a 32×32 array, but still have a limited field of view (FOV). Our research aims to maintain or lower the element count of matrix arrays while increasing the FOV by increasing the element size of the arrays and use advanced beamformers to maximize quality. Elements were coupled together by summing channel data to imitate element sizes up to eight wavelengths. The delay-and-sum (DAS), null subtraction imaging (NSI), phase coherence factor (PCF), and sign coherence factor (SCF) beamformers were compared. The NSI beamformer eliminated grating lobes in simulated point-spread functions of large-element matrix arrays. The contrast of cyst phantom images increased from 15 dB with DAS to 19 dB using NSI. Spatial resolution also improved in wire phantoms from 80 μm with DAS to 60 μm using NSI. Our results suggest that matrix arrays could be built with larger FOV and lower element counts with little to no loss to image quality.