Key result
GTT occludes primarily via fibrin-rich coagulation rather than high-shear arterial platelet aggregation.
Why the study?
It was unclear whether thrombosis in the global thrombosis test point-of-care device is dominated by coagulation-triggered red clot or high shear-induced white clot.
Does the Global Thrombosis Test (GTT) occlude primarily due to coagulation-triggered red clot or high shear-induced white clot?
Does the Global Thrombosis Test (GTT) occlude primarily due to coagulation-triggered red clot or high shear-induced white clot?
The Global Thrombosis Test occludes primarily due to coagulation rather than high shear-induced platelet aggregation, challenging its utility as a specific measure of arterial thrombosis.
Questions GTT specificity for arterial thrombosis; leaves open its clinical role pending human validation.
The global thrombosis test (GTT) is a point of care device that tests thrombotic and thrombolytic status. The device exposes whole blood flow to a combination of both high and low shear stress past and between ball bearings potentially causing thrombin and fibrin formation. The question arises as to whether thrombosis in the GTT is dominated by coagulation-triggered red clot or high shear-induced white clot. We investigated the nature of the thrombus formed in the GTT, the device efficacy, human factors use, and limitations. The GTT formed clots that were histologically fibrin-rich with trapped red blood cells. The occlusion time (OT) was more consistent with coagulation than high shear white clot and was strongly lengthened by heparin and citrate, two common anticoagulants. The clot was lysed by tissue plasminogen activator (tPA), also consistent with a fibrin-rich red clot. Changing the bead to a collagen-coated surface and eliminating the low shear zone between the beads induced a rapid OT consistent with a platelet-rich thrombus that was relatively resistant to heparin or tPA. The evidence points to the GTT as occluding primarily due to fibrin-rich red clot from coagulation rather than high shear platelet aggregation and occlusion associated with arterial thrombosis.
No takes yet. Share an insight, caveat, or question.
Clavería et al. (2021) studied Thrombosis. Global thrombosis test (GTT) was evaluated on Nature of the thrombus formed in the GTT. The global thrombosis test occludes primarily due to fibrin-rich red clot from coagulation rather than high shear platelet aggregation associated with arterial thrombosis.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: