Key result
Edwards Commander delivery system insertion expands 14 F and 16 F eSheath outer diameters during TAVR.
Why the study?
What are the exact changes in the outer diameters of 14 F and 16 F eSheaths during the insertion of the Edwards Commander Delivery System and after TAVR?
Observational
What are the exact changes in the outer diameters of 14 F and 16 F eSheaths during the insertion of the Edwards Commander Delivery System and after TAVR?
Answers are grounded in this paper and cite their sources.
The outer diameters of 14 F and 16 F eSheaths significantly increase during the advancement of the delivery system and after TAVR, exceeding their nominal sizes.
Innovative catheter systems with lower-profile sheaths and a dynamic expansion mechanism (DEM) were recently introduced for transcatheter aortic valve replacement (TAVR). However, the labeling of 14 F and 16 F eSheaths denote the inner nominal diameter. Exact changes of the clinically relevant outer diameters during usage are not available. eSheaths were measured every 30 mm using a digital caliper. Unused 14 F and 16 F eSheaths served as controls. Maximum eSheath diameters were measured after insertion of the Edwards Commander Delivery System (ECDS) into 14 F and 16 F eSheaths.Finally, eSheaths were retrieved and measured after TAVR. Outer diameters of control 14 F eSheaths were 5.8 mm and 6.50 mm for the 16 F eSheath. Introduction of the 23 mm and 26 mm ECDS into 14 F eSheaths showed a maximum diameter of 7.65 mm and 7.64 mm (P = NS). Introduction of the 29 mm ECDS into the 16 F eSheath showed the greatest diameter of 8.18 mm (P = 0.03). After TAVR, diameters of the 14 F eSheaths were 7.14 mm (23 mm valve) and 7.26 mm (26 mm valve) (P = NS), while 16 F eSheaths were 8.10 mm (29 mm valve) (P ≤ 0.03). Nominal 14 F and 16 F eSheaths showed a significant increase of the outer diameter during advancement of the ECDS and after TAVR implantation.
No takes yet. Share an insight, caveat, or question.
Koehler et al. (2015) conducted an observational in Transcatheter aortic valve replacement. Edwards Commander Delivery System insertion vs. Unused 14 F and 16 F eSheaths was evaluated on Maximum outer diameter of eSheaths (p=0.03). Insertion of the Edwards Commander Delivery System significantly increased the outer diameter of nominal 14 F and 16 F eSheaths during advancement and after TAVR implantation.
Synapse has enriched 10 related papers. Consider them for comparative context: