Randomized trial shows enhanced detection of cardiac troponin I in serum, suggesting improved CVD diagnostics.

Key Points

  • This study aims to develop a sensitive biosensor for detecting cardiac troponin I (cTnI) as a potential diagnostic tool for cardiovascular disease.
  • Developed a biosensor using silane-modified zeolite on interdigitated electrodes (IDE).
  • Attached glutaraldehyde premixed with anti-cTnI to the IDE to enhance antibody attachment.
  • Tested the sensor's ability to detect cTnI at concentrations ranging from 1 fM to 1 nM.
  • Detected cTnI at concentrations as low as 1 fM with a linear regression curve (1 fM to 1 nM) and R² = 0.9417.
  • Current increase observed with cTnI-spiked serum, confirming sensitivity.
  • Specificity confirmed through unchanged current responses with other proteins and nonimmune antibodies.