Why the study?
This study evaluated the accuracy of the Omron HEM-7600T automated oscillometric upper-arm blood pressure monitor according to the ANSI/AAMI/ISO 81060-2:2013 protocol.
Does the Omron HEM-7600T automated oscillometric upper-arm blood pressure monitor accurately measure blood pressure compared to a mercury sphygmomanometer in the general population without arrhythmias?
Does the Omron HEM-7600T automated oscillometric upper-arm blood pressure monitor accurately measure blood pressure compared to a mercury sphygmomanometer in the general population without arrhythmias?
The Omron HEM-7600T automated blood pressure monitor meets ANSI/AAMI/ISO validation criteria and is accurate for home blood pressure measurement in adults without arrhythmias.
Answers are grounded in this paper and cite their sources.
Supports accuracy for home BP monitoring in selected adults; leaves open outcome validation in broader populations.
OBJECTIVE: The present study aimed to evaluate the accuracy of the Omron HEM-7600T automated oscillometric upper-arm blood pressure (BP) monitor according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 protocol. METHODS: We recruited individuals until we obtained 86 eligible participants whose BP levels and ranges of arm circumference met the BP distribution requirements specified by the ANSI/AAMI/ISO. In each individual, we sequentially measured systolic and diastolic blood pressures (DBPs) using a mercury sphygmomanometer (two observers) and the HEM-7600T device (one supervisor) in the left upper arm. Data analysis was performed according to the ANSI/AAMI/ISO guidelines. RESULTS: A total of 101 individuals were recruited, and 86 were finally included in the analysis. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP was -0.6 ± 6.6/1.0 ± 6.5 mmHg (systolic/diastolic). The data were in accordance with criterion 1 of the ANSI/AAMI/ISO 81060-2:2013 standard requirements (≤ 5 ± ≤ 8 mmHg). In addition, the mean device-observer difference in the 86 participants was -0.6 ± 5.65 mmHg for systolic blood pressure and 1.0 ± 5.96 mmHg for DBP, fulfilling criterion 2 with an SD of ≤6.91 mmHg for systolic blood pressure and ≤6.87 mmHg for DBP. CONCLUSION: The Omron HEM-7600T BP monitor fulfilled the requirements of the ANSI/AAMI/ISO validation protocol in adults and can be recommended for home BP measurements.