Why the study?
This study evaluated the accuracy of the Omron HEM-7600T automated oscillometric upper-arm blood pressure monitor according to the ANSI/AAMI/ISO 81060-2:2013 protocol.
Does the Omron HEM-7600T automated oscillometric upper-arm blood pressure monitor accurately measure blood pressure compared to a mercury sphygmomanometer in the general population without arrhythmias?
Population
86 eligible participants meeting ANSI/AAMI/ISO blood pressure and arm circumference requirements
Comparison
Omron HEM-7600T device vs mercury sphygmomanometer
Design
Device validation study
Authors
Loading...
Supports accuracy for home BP monitoring in selected adults; leaves open outcome validation in broader populations.
Does the Omron HEM-7600T automated oscillometric upper-arm blood pressure monitor accurately measure blood pressure compared to a mercury sphygmomanometer in the general population without arrhythmias?
The Omron HEM-7600T automated blood pressure monitor meets ANSI/AAMI/ISO validation criteria and is accurate for home blood pressure measurement in adults without arrhythmias.