Answers are grounded in this paper and cite their sources.
Cochlear implants usually employ an automatic gain control (AGC) system as a first stage of processing. AGC1 was a fast-acting (syllabic) compressor. AGC2 was a dual-time-constant system; it usually performed as a slow-acting compressor, but incorporated an additional fast-acting system to provide protection from sudden increases in sound level. Six experienced cochlear-implant users were tested in a counterbalanced order, receiving one-month of experience with a given AGC type before switching to the other type. Performance was evaluated shortly after provision of a given AGC type and after one-month of experience with that AGC type. Questionnaires, mainly relating to listening in quiet situations, did not reveal significant differences between the two AGC types. However, fixed-level and roving-level tests of sentence identification in noise both revealed significantly better performance for AGC2. It is suggested that the poorer performance for AGC1 occurred because AGC1 introduced cross-modulation between the target speech and background noise, which made perceptual separation of the target and background more difficult.
No takes yet. Share an insight, caveat, or question.
Boyle et al. (2009) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: