) chelates, and finally, time-resolved fluorometry was used to measure the fluorescence values. We successfully established a TRFIA method for the detection of human COVID-19 total antibodies, and the cutoff value was 2.02. There was no cross-reactivity with the negative reference of the National Reference Panel for IgM and IgG antibodies to COVID-19. The CV of the precision assay was 3.19%, and the assay could be stored stably for 15 days at 37°C. Compared with that of the colloidal gold method and chemiluminescence method, the sensitivity of the TRFIA method was higher, and the false positive/negative rate was lower. This established TRFIA has high sensitivity, accuracy, and specificity, which indicates that this method provides a new detection method for the high-throughput routine diagnosis of COVID-19.