A high-resolution vibrational electron energy-loss spectrometer of 127°-cylindrical electrostatic deflection type has been constructed for surface studies. In the direct mode, an electron current at the collector entrance of 2.3×10−11 A is obtained with the energy spread (full width at half-maximum) of 60 cm−1 for electrons whose energy at the sample position is 5 eV. A ghost reduction slit is used to suppress spurious peaks due to the electron reflection from electrodes of the analyzer. Electron energy-loss spectroscopy can be combined in situ with low-energy electron diffraction, Auger electron spectroscopy, and thermal-desorption mass spectroscopy. Design and performance of the apparatus are reported. The vibrational spectrum of the Ni(110) (1×2)-H surface formed at room temperature is presented, and the location of hydrogen atoms on the surface is discussed.
No takes yet. Share an insight, caveat, or question.
Nishijima et al. (1982) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: