Key result
A commercially available AED demonstrated 98% sensitivity and 100% specificity for recognizing ventricular fibrillation in a swine model of commotio cordis, successfully terminating all episodes.
Why the study?
Does an automated external defibrillator accurately detect and terminate ventricular fibrillation in a juvenile swine model of commotio cordis?
Population
26 anesthetized juvenile swine undergoing 50 ventricular fibrillation inductions via precordial impact from…
Comparison
Automated external defibrillator applied after… vs Different time intervals of ventricular…
Design
Preclinical, Animals were randomized to defibrillation after 1, 2, 4, or 6…
Authors
Loading...
AED accuracy in swine commotio cordis is encouraging; leaves open translation to pediatric athletes pending human data.
RCT (n=26)
1, 2, 4, or 6 minutes of ventricular fibrillation
Does an automated external defibrillator accurately detect and terminate ventricular fibrillation in a juvenile swine model of commotio cordis?
In a juvenile swine model of commotio cordis, a commercially available AED accurately detected and successfully terminated ventricular fibrillation, supporting its potential life-saving use in children on athletic fields.
Link et al. (2003) conducted an RCT in Cardiac arrest due to commotio cordis (n=26). Automated external defibrillator (AED) vs. Different time intervals (1, 2, 4, or 6 minutes) was evaluated on Sensitivity for recognition of ventricular fibrillation. A commercially available AED demonstrated 98% sensitivity and 100% specificity for recognizing ventricular fibrillation in a swine model of commotio cordis, successfully terminating all episodes.