Heterophile antibodies linked to falsely elevated hs-TnI on the Abbott Alinity i assay.
Case Report (n=1)
No
Heterophile antibodies can cause spuriously elevated high-sensitivity troponin I on specific assays, highlighting the need for clinical correlation and alternate assay testing when laboratory results do not match the clinical picture.
Answers are grounded in this paper and cite their sources.
Clinicians should consider heterophile interference in discrepant hs-TnI results on Abbott Alinity i; this case leaves open its prevalence and assay impact.
Hereby we describe a case of a 59-year-old female patient with persistently elevated high-sensitivity troponin I (hs-TnI) over the course of almost four years measured on Alinity i with the corresponding assay (Abbott Laboratories, Chicago, USA). The patient underwent multiple extensive cardiological evaluations, but none of them suggested acute or chronic cardiac damage. Therefore, interference in measurement was suspected by the attending cardiologist and a detailed, stepwise laboratory investigation was undertaken in the sample with initial hs-TnI result of 2077 ng/L. Serial sample dilutions (1:2,1:5,1:10) did not match the expected, calculated hs-TnI concentrations, yielding both huge positive biases (62, 109 and 139%, respectively) and absolute differences (639, 453 and 290 ng/L, respectively). Precipitation with polyethylene-glycol, pretreatment in heterophilic blocking tubes (HBT) and immunoglobulin G depletion yielded hs-TnI results below the assay's diagnostic cut-off (< 15.6 ng/L). Alternate hs-TnI immunoassays (Siemens Healthineers, Beckman Coulter and Snibe) and measurement with the high-sensitivity troponin T (hs-TnT) assay yielded results below assays' specific cut-off values. This investigation confirmed that results of hs-TnI obtained by the Abbott assay were spuriously elevated. Significant lowering of hs-TnI after HBT pretreatment indicated that heterophile antibodies are the most probable source of interference. Based on this finding, it was entered in the patient's medical record that future determinations of cardiac troponin should be performed with an alternate hs-TnI or hs-TnT assay. This case emphasizes that analytical interferences are usually immunoassay-dependent. Evaluation of laboratory results in the clinical context and close collaboration between laboratory and clinical staff is crucial for their recognition.
No takes yet. Share an insight, caveat, or question.
Đerek et al. (2025) conducted a case report in Falsely elevated high-sensitivity troponin I (n=1). Laboratory investigation for analytical interference was evaluated on hs-TnI concentration after pretreatment. Detailed laboratory investigation confirmed that persistently elevated high-sensitivity troponin I results on the Abbott Alinity i assay were falsely elevated due to heterophile antibodies.
Synapse has enriched 10 related papers. Consider them for comparative context: