Key result
Echocardiography-guided temporary pacing reduced time to pacing (439 vs 716 s; P<0.0001) and overall complications (OR 0.18; 95% CI 0.06-0.49; P<0.001) compared to fluoroscopy guidance.
Why the study?
Does echocardiography-guided temporary pacing improve time to pacing and reduce complications compared to fluoroscopy-guided pacing in patients needing urgent pacing?
Observational (n=106)
No
Does echocardiography-guided temporary pacing improve time to pacing and reduce complications compared to fluoroscopy-guided pacing in patients needing urgent pacing?
Effect estimate: OR 0.18 (95% CI 0.06-0.49)
Absolute Event Rate: 11.3% vs 41.5%
p-value: p=<0.001
Echocardiography-guided temporary pacing via the internal jugular vein is faster and associated with fewer complications than standard fluoroscopy-guided pacing via the femoral vein in emergency settings.
May support echocardiography guidance for urgent temporary pacing; leaves open need for randomized confirmation.
BACKGROUND: The insertion of a temporary pacemaker can be a lifesaving procedure in the emergency setting. OBJECTIVES: This is an observational monocentric study comparing echocardiography-guided temporary pacemaker via the right internal jugular vein to standard fluoroscopy-guided temporary pacemaker via the femoral vein; the procedure was tested for noninferiority. METHODS: Patients needing urgent pacing were consecutively enrolled. Primary efficacy endpoints were time to pacing and need for catheter replacement. Primary safety endpoint was a composite outcome of overall complications. RESULTS: One hundred and six patients (77 ± 10 years) were enrolled: 53 underwent echocardiographic-guided and 53 fluoroscopy-guided temporary pacemaker. Baseline characteristics of the two groups of treatment were similar. Time to pacing was shorter in the echocardiography-guided than in the fluoroscopy-guided group (439 ± 179 vs. 716 ± 235 s; P<0.0001; power 100%). During the pacing (54 ± 35 h), there was a higher incidence of pacemaker malfunction in the fluoroscopy-guided group [15 vs. 3 patients; odds ratio (OR) 6.5, confidence interval (CI) 95% 1.9-29.7, P<0.001; power 5.7%] and there was a significantly lower incidence of complications in the echocardiography-guided temporary pacemaker group (6 vs. 22 patients; OR 0.18, CI 95% 0.06-0.49, P<0.001; echocardiography-guided temporary pacemaker events rate 0.1929 vs. fluoroscopy-guided temporary pacemaker events rate 1.398 per 100 person-hours paced, P<0.0001). In the standard group there was one death attributable to a temporary pacemaker complication (sepsis). CONCLUSION: Echocardiography-guided temporary pacemaker is a well-tolerated procedure that could allow reliable insertion of a temporary pacemaker; therefore, it is a well-tolerated option in an emergency setting and in hospitals where fluoroscopy is not available.
No takes yet. Share an insight, caveat, or question.
Pinneri et al. (2012) conducted an observational in Urgent pacing requirement (n=106). Echocardiography-guided temporary pacemaker vs. Fluoroscopy-guided temporary pacemaker was evaluated on Composite outcome of overall complications (OR 0.18, 95% CI 0.06-0.49, p=<0.001). Echocardiography-guided temporary pacing reduced time to pacing (439 vs 716 s; P<0.0001) and overall complications (OR 0.18; 95% CI 0.06-0.49; P<0.001) compared to fluoroscopy guidance.