An enzyme-linked immunosorbent assay (ELISA) for the measurement of human tissue-type plasminogen activator (t-PA) was developed. Microtiter plates were coated with a mixture of two monoclonal antibodies and bound t-PA was quantitated with a third monoclonal antibody linked to peroxidase. The lower limit of sensitivity of the assay was 0.2 ng of t-PA per ml. The concentration of antigen in citrated plasma of human subjects was found to be 3.4 +/- 0.8 ng/ml. The assay had a good reproducibility with values of 3.8, 6.5 and 4.9 percent respectively for the intra-, inter-assay and inter-dilution variation coefficients. The results of the ELISA assay on plasma samples from patients during thrombolytic therapy with t-PA correlated very well, over a wide concentration range, with those obtained with a previously described two-site immuno-radiometric assay (r = 0.96). This ELISA with monoclonal antibodies constitutes a stable and reproducible set of reagents for the measurement of t-PA antigen in biological fluids, avoiding the disadvantages of the use of radioisotopes and of polyclonal antibodies.
No takes yet. Share an insight, caveat, or question.
Holvoet et al. (1985) studied this question.