Key result
3D-multiplane planimetry yields larger AVA than standard continuity equations but aligns below 1.5 cm2.
Why the study?
The study aimed to evaluate the reliability and consistency of measuring aortic valve area using 3D transesophageal echocardiography compared with continuity equation methods.
Does 3D-multiplane reconstruction planimetry reliably measure aortic valve area compared to established continuity equation methods in patients undergoing transesophageal echocardiography?
Cross-Sectional (n=119)
Blinded readers
Does 3D-multiplane reconstruction planimetry reliably measure aortic valve area compared to established continuity equation methods in patients undergoing transesophageal echocardiography?
3D transesophageal echocardiography planimetry provides reliable aortic valve area measurements comparable to established continuity equation methods, particularly in patients with more severe stenosis.
Answers are grounded in this paper and cite their sources.
3D planimetry may overestimate AVA versus continuity equation except in severe cases; leaves open validation for TEE grading accuracy.
We aim to evaluate the reliability and consistency of measuring the aortic valve area (AVA) using 3-dimensional (3D) transesophageal echocardiography and compare it with invasive and noninvasive methods using a continuity equation (CE). Measurements were taken from 119 patients with different severity of aortic stenosis and with normal aortic valve who underwent elective transesophageal echocardiography encompassing the whole spectrum of aortic opening. Three methods were compared to determine AVA. First, the effective AVA was calculated with the standard CE, where the left ventricular outflow tract area was calculated from its 2-dimensional diameter (AVA-CEstd). Second, a modified CE method (AVA-CEmod) was used, in which the left ventricular outflow tract area was measured using 3D-multiplane reconstruction. Third, the geometric AVA was directly measured using 3D-multiplane reconstruction planimetry (AVA-3D). Interobserver and intraobserver variability were analyzed using intraclass correlation coefficients (ICCs). The values were measured by two blinded readers for interobserver variability and by one observer on the same dataset. AVA-3D was significantly larger than AVA-CEmod and AVA-CEstd (1.87 ± 1.00 cm2 vs 1.81 ± 0.92 cm2 p = 0.03 and 1.87 ± 1.00 cm2 vs 1.71 ± 0.85 cm2 p <0.001). However, in the subset of patients with AVA-3D <1.5 cm2, there was no significant difference between AVA-3D and AVA-CEmod (1.06 ± 0.24 vs 1.08 ± 0.26 cm2, paired t test: t = 0.77, degree of freedom = 58, p = 0.44). The ICC between the measurements of AVA-3D and AVA-CEmod (ICC 0.979), and AVA-3D and AVA- CEstd (ICC 0.940), were excellent. AVA-3D delivers very similar results as compared with more established echocardiographic parameters. The difference between effective and geometric AVA did not appear to be clinically relevant in patients with a higher degree of stenosis. We aim to evaluate the reliability and consistency of measuring the aortic valve area (AVA) using 3-dimensional (3D) transesophageal echocardiography and compare it with invasive and noninvasive methods using a continuity equation (CE). Measurements were taken from 119 patients with different severity of aortic stenosis and with normal aortic valve who underwent elective transesophageal echocardiography encompassing the whole spectrum of aortic opening. Three methods were compared to determine AVA. First, the effective AVA was calculated with the standard CE, where the left ventricular outflow tract area was calculated from its 2-dimensional diameter (AVA-CEstd). Second, a modified CE method (AVA-CEmod) was used, in which the left ventricular outflow tract area was measured using 3D-multiplane reconstruction. Third, the geometric AVA was directly measured using 3D-multiplane reconstruction planimetry (AVA-3D). Interobserver and intraobserver variability were analyzed using intraclass correlation coefficients (ICCs). The values were measured by two blinded readers for interobserver variability and by one observer on the same dataset. AVA-3D was significantly larger than AVA-CEmod and AVA-CEstd (1.87 ± 1.00 cm2 vs 1.81 ± 0.92 cm2 p = 0.03 and 1.87 ± 1.00 cm2 vs 1.71 ± 0.85 cm2 p <0.001). However, in the subset of patients with AVA-3D <1.5 cm2, there was no significant difference between AVA-3D and AVA-CEmod (1.06 ± 0.24 vs 1.08 ± 0.26 cm2, paired t test: t = 0.77, degree of freedom = 58, p = 0.44). The ICC between the measurements of AVA-3D and AVA-CEmod (ICC 0.979), and AVA-3D and AVA- CEstd (ICC 0.940), were excellent. AVA-3D delivers very similar results as compared with more established echocardiographic parameters. The difference between effective and geometric AVA did not appear to be clinically relevant in patients with a higher degree of stenosis.
No takes yet. Share an insight, caveat, or question.
Blaser et al. (2023) conducted a cross-sectional in Aortic stenosis (n=119). 3D-multiplane reconstruction planimetry (AVA-3D) vs. Standard continuity equation (AVA-CEstd) was evaluated on Aortic valve area (AVA) in cm2 (p=<0.001). 3D-multiplane reconstruction planimetry yielded larger aortic valve areas than standard continuity equations, but results were similar in patients with AVA <1.5 cm2.
Synapse has enriched 10 related papers. Consider them for comparative context: