This work presents a new steerable needle to facilitate active steering toward predefined target locations. It focuses on mechanical aspects and design choices in relation to the observed response in a tissue phantom. Tip steering with two rotational degrees of freedom was achieved by a tendon actuated ball joint mechanism. During insertion, the flexible cannula bends as a result of asymmetric tip-tissue interaction forces. The stylet was equipped with fiber Bragg gratings to measure the needle shape and tip position during use. A PI-controller was implemented to facilitate steering to predefined targets. During the validation study, nine targets were defined at a depth of 100 mm below the gelatin surface. One was located below the insertion point, the others at a radial offset of 30 mm in each of the eight principle steering directions. Per location, six repetitions were performed. The targeting accuracy was 6.2 ± 1.4 mm (mean ± std). The steering precision was 2.6 ± 1.1 mm. The ability to steer with this new needle steering approach is presented and the mechanical characteristics are discussed for this representative subset of steering directions.