A technique for steering of flexible bevel-tipped needles through tissue, providing proportional control of trajectory curvature by means of duty-cycled rotation or spinning during insertion, has been presented previously, and tested in vitro in gelatin samples. The present paper presents the results of testing under more authentic conditions. Thirty-two needle insertions were performed in cadaver brain tissue. Needle insertion was controlled interactively by the surgeon, adjusting parameters intraoperatively as needed based on fluoroscope images acquired at intervals of 1-2 cm of insertion depth. The average target acquisition error was 1.80 +/- 1.33 mm.
No takes yet. Share an insight, caveat, or question.
Minhas et al. (2009) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: