Comparative evaluation reveals variable glomerular filtration rate estimates across cystatin C reagent and equation combinations, indicating low-concentration discrepancies challenge standardization.
Background: Serum cystatin C (cysC), which is less affected by sex, race, and muscle mass than creatinine, is a useful biomarker of the estimated glomerular filtration rate (eGFR). The standardization of cysC measurements remains controversial, although a certified reference material (ERM-DA471/IFCC) is available. Moreover, the effect of combinations of cysC reagents and equations for eGFR is unclear. Methods: ). Results: . Regression analysis showed concordance between the reagents within 0.85-4.40 mg/L when using ±7.61% total allowable error. Lin's concordance correlation coefficient of eGFR, by combining the measuring system and equation, varied from 0.73 to 1.00. Conclusions: The equivalence of cysC values at low concentrations (<0.85 mg/L) between the two reagents was unsatisfactory. Results obtained with different measurement systems could lead to larger differences in eGFR varying with the combination.