December 1, 2001Blood Pressure Monitoring174 citations

Oscillometric blood pressure measurement: progress and problems

View Full Paper
GMGert A. van Montfrans

Key Result

Oscillometric blood pressure measurement devices can deviate by >5 mmHg in individual subjects despite high ratings, highlighting the need for improved algorithms and simulators.

Structured PICO

I
Intervention
Oscillometric blood pressure measurement devices

Despite passing validation criteria, oscillometric blood pressure devices can have clinically relevant deviations, highlighting the need for better validation protocols and more robust algorithms.

Limitations

  • The currently employed maximal amplitude algorithm has many drawbacks as parameter identification points depend on factors like pulse pressure, heart rate, and arterial stiffness.
  • Modern pattern recognition algorithms have not yet produced convincing results.

Abstract

Oscillometric blood pressure measurement has become very popular, but although a number of devices have now passed both the Association for the Advancement of Medical Instrumentation and British Hypertension Society criteria, complacency with the state of the technique is as yet premature. In individual subjects, a substantial number of readings may deviate more than a clinically relevant 5 mmHg in devices that have earned a British Hypertension Society grade A rating. The marketing of pressure-wave-simulating devices is a welcome development as monitors can now be tested for reproducibility; an intra-device standard deviation of less than 2 mmHg has been proposed as the limit. Authors suggest that these simulators are currently better suited to intra- than between-device testing since they are not yet fully confident that the simulated waveforms are indistinguishable from the man-made pressure waves. Simulators should, however, be incorporated into our standard validation protocols in order eventually to obviate the human, fallible, factor in the validation protocols. The currently employed maximal amplitude algorithm has many drawbacks as the parameter identification points for systolic and diastolic pressure depend on many factors, for example pulse pressure, heart rate and arterial stiffness. These errors have now been demonstrated in clinical studies. Modern pattern recognition algorithms are being constructed but have not yet produced convincing results. As repeatedly stated, the development of a more robust and more widely applicable algorithm than the maximal amplitude approach should be allocated a high priority.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Gert A. van Montfrans (2001) conducted a review in Blood pressure measurement. Oscillometric blood pressure measurement was evaluated. Oscillometric blood pressure measurement devices can deviate by >5 mmHg in individual subjects despite high ratings, highlighting the need for improved algorithms and simulators.

synapsesocial.com/papers/6a09e0663f847eb6b633f19ehttps://doi.org/10.1097/00126097-200112000-00004
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Indirect Measurement of Eye-Level Arterial Pressure by the Volume Oscillometric Method1990 · 4 citations
  2. 2Effect of Age on Brachial Artery Wall Properties Differs From the Aorta and Is Gender Dependent2000 · 386 citations
  3. 3A simulation-based evaluation of nine oscillometric blood pressure monitors for self-measurement2000 · 6 citations
  4. 4A Consensus View on the Technique of Ambulatory Blood Pressure Monitoring1995 · 118 citations
  5. 5Mathematical Modeling of Noninvasive Blood Pressure Estimation Techniques—Part II: Brachial Hemodynamics1995 · 26 citations